Instruction Manual - TORBAL AGC SeriesCHAPTER 6.WEIGHING1. To begin weighing, press the power button ( ) to turn the scale on. The scale will initia
Instruction Manual - TORBAL AGC Series4. When it has finished re-zeroing, the scale will return to weighing mode and the display will indicate 0.000g.
Instruction Manual - TORBAL AGC Series6.1.3.Clearing a Tare1. To clear the tare, remove the tared object along with the net weight from the pan. The s
Instruction Manual - TORBAL AGC SeriesCHAPTER 7.MAIN MENUMain Menu is used to configure the scale and its weighing modes. Before entering the menu, c
Instruction Manual - TORBAL AGC Series7.1.1.1.Start (Initializing Automatic Calibration)To calibrate the scale, follow the steps below:1. Press “MENU”
Instruction Manual - TORBAL AGC Series5. Before calibration begins, the scale will perform automatic tare. TARING will be displayed during auto tare.6
Instruction Manual - TORBAL AGC Series2. When CALIB is displayed, press “T” (YES).3. The scale will display commands Start and Setup sequentially. Pre
Instruction Manual - TORBAL AGC Series7.1.1.2.1. Calibration Adjustment The Calibration Adjustment function allows the internal calibration mass value
Instruction Manual - TORBAL AGC SeriesTo Change To Accept5. Command LOAD will be displayed. At any time during the LOAD command, place a single calibr
Instruction Manual - TORBAL AGC Series8. Once the weight is removed the scale will continue with calibration adjustment. CAL will be displayed during
Instruction Manuala Important handling Cautions and Warnings Always handle your scale with care. Damage caused by improper handling is not covered u
Instruction Manual - TORBAL AGC Series3. A dash will appear on the display indicating to enter a desired time interval. The time interval can be set b
Instruction Manual - TORBAL AGC Series1. Press the “T” (YES) key when “TEMP” is displayed in the calibration setup menu. 2. “°C - 0” and the current t
Instruction Manual - TORBAL AGC Series5. When the desired setting is displayed press the “Function” key to accept and save the selection. The scale wi
Instruction Manual - TORBAL AGC Series2. When command UNIT is displayed press the “T” (Yes) key.3. The scale will display available units of measure
Instruction Manual - TORBAL AGC SeriesImportant: - The symbols for grams (g), kilograms (kg), carats (ct), pounds (lb), and Newton (n) will appear to
Instruction Manual - TORBAL AGC Series1. Enter the Main Menu by pressing the “MENU” key.2. When MODE is displayed press the “T” (YES) key.3. The appli
Instruction Manual - TORBAL AGC Series2. Commands An On and An OFF will be displayed sequentially. To enable animal weighing and proceed with configur
Instruction Manual - TORBAL AGC Series4. Once the desired operating mode for animal weighing has been selected, the scale will display time intervals
Instruction Manual - TORBAL AGC Series(B) Parts Counting1. To enable parts counting press the “T” (YES) key when Parts is displayed.28Table 7.1.3.A: A
Instruction Manual - TORBAL AGC Series2. Commands Co On and Co OFF will be displayed sequentially. To enable parts counting and proceed with configura
Instruction ManualCCAUTIONARY NOTES AND PRECAUTIONSThe TORBAL Precision scale is a sensitive and a delicate instrument. Always handl
Instruction Manual - TORBAL AGC SeriesNote: Larger sample size results in a more accurate average piece weight. Select a sample size greater than 10 p
Instruction Manual - TORBAL AGC Series3. After enabling check weighing, DONE will be displayed indicating that the function has been configured and th
Instruction Manual - TORBAL AGC Series3. Display hold can operate in three different operating modes: automatic, semi-automatic and manual (see table
Instruction Manual - TORBAL AGC Series(E) Percent Weighing1. To enable percent weighing, press the “T” (YES) key when Percent is displayed33Table 7.1.
Instruction Manual - TORBAL AGC Series2. Commands Pe On and Pe OFF will be displayed sequentially. To enable percent weighing press the “T” (YES) key
Instruction Manual - TORBAL AGC Series2. Commands tot On and tot OFF will be displayed sequentially. To enable totalizing and proceed with configurati
Instruction Manual - TORBAL AGC Series7.1.4.Automatic Zero Setting Mechanism (AZSM)All AGC scales are equipped with AZSM, the “Auto
Instruction Manual - TORBAL AGC Series3. Command AUT On and AUT OFF will flash. 4. To enable AZSM press the “T” (YES) key when AUT On is displayed. To
Instruction Manual - TORBAL AGC Series7.1.5.GLP Print and Data ConfigurationWhen enabled, GLP data will appear on every printed transaction receipt. T
Instruction Manual - TORBAL AGC Series2. When command GLP is displayed press the “T” (YES) key.3. Commands Set, Model, Unit ID, User ID, and Project I
Instruction Manual - TORBAL AGC SeriesCHAPTER 1.SPECIFICATION4ModelAGC100 AGC200 AGC300 AGC500 AGC1000 AGC2000 AGC3000 AGC4000Capacity100g 200g 300g
Instruction Manual - TORBAL AGC Series6. Once enabled, GLP options will begin to scroll on the display: Model No., Unit ID, User ID, and Project No. 7
Instruction Manual - TORBAL AGC Series9. To enter an ID or a number, use the following keys: 0(NO) to increment a digit, “T” (YES) to accept and go
Instruction Manual - TORBAL AGC Series7.1.6.Printing Data and RS232 Communication Port ConfigurationsThe printing function is used to print data via t
Instruction Manual - TORBAL AGC Series4. To configure the Baud rate settings press the “T” key when Baud is displayed. To configure the Parity setting
Instruction Manual - TORBAL AGC Series6. After pressing “T” when Parity is displayed, the scale will display available parity settings: none, even, od
Instruction Manual - TORBAL AGC Series7.1.6.2. Torbal Printer Configuration (RXP-4))To connect the AGC scale to a TORBAL (RXP-4) printer, use the foll
Instruction Manual - TORBAL AGC Series7.1.6.3. Printout ReceiptsAfter completing a weighing transaction, result data receipt may be printed. To initia
Instruction Manual - TORBAL AGC Series7.1.7.Time and DateTo set the current time and date, follow the steps below:1. Enter the Main Menu by pressing t
Instruction Manual - TORBAL AGC Series3. Commands Time and Date will flash. Time:1. To set the time press the “T” (YES) key when Time is displayed.2.
Instruction Manual - TORBAL AGC SeriesDate:1. To set the date press “T” (YES) when Date is displayed.2. The current date setting will be displayed. To
Instruction Manual - TORBAL AGC SeriesKEYS AND DISPLAY INDICATORSKey Primary Function Secondary FunctionPower On and Power Off -T (YES)Tare – used to
Instruction Manual - TORBAL AGC Series2. When RESET is displayed press the “T” (YES) key.3. The scale will display YES – NO in order to confirm reset
Instruction Manual - TORBAL AGC SeriesTable 7.1.5 Factory Default SettingsWeighing Function: Always enabledAnimal Weighing Function: DisabledFiltering
Instruction Manual - TORBAL AGC SeriesCHAPTER 8.FUNCTION MENU: USING THE MODES AND FUNCTIONSAll enabled and configured modes as well as applications
Instruction Manual - TORBAL AGC Series(A) Animal Weighing1. Before weighing the animal, place an animal weighing container on the scale’s pan and pres
Instruction Manual - TORBAL AGC SeriesNote: In Automatic and Semi-Automatic modes the minimum weighing mass must be equal to or greater than 10d. To s
Instruction Manual - TORBAL AGC Series4. After the time interval has elapsed the result will be displayed. The animal can now be removed from the pan.
Instruction Manual - TORBAL AGC Series(B) Parts Counting1. Place an empty container on the pan and press the “T” key to tare.2. After taring the conta
Instruction Manual - TORBAL AGC Series3. After confirming the sample size selection, the display will read SAPL On, indicating that you may now place
Instruction Manual - TORBAL AGC Series6. Once the weight of the sample stabilizes and the stability indicator appears on the display, press “T” to con
Instruction Manual - TORBAL AGC Series9. Once an accurate piece count has been taken, the container and its contents may be removed from the scale. Af
Instruction Manual - TORBAL AGC SeriesCHAPTER 3.COMMANDS AND ABBREVIATIONSCommand or Abbreviation Description- - - - - -Taring or re-zeroing in progr
Instruction Manual - TORBAL AGC Series(C) Check WeighingThe check weighing function is used to compare a weight of an object against a preset target
Instruction Manual - TORBAL AGC Series4. To enter a weight, use the following keys: the 0(NO) key to increment a digit; “T” (YES) to accept and go t
Instruction Manual - TORBAL AGC Series7. To change the OVER limit follow the directions in Step 4 of this section and Table 8.C.4.8. Once both limits
Instruction Manual - TORBAL AGC Series(D) Display HoldThe Display Hold function is used to retain a weighing result static on the LCD display after it
Instruction Manual - TORBAL AGC Series3. WeighingAutomatic and Semi-Automatic:When set to Auto and Semi-Auto after weight has been added to the pan,
Instruction Manual - TORBAL AGC SeriesSemi-Automatic and Manual:When set to Semi-Automatic or Manual, the weighing result will remain displayed and it
Instruction Manual - TORBAL AGC Series2. Press “T” to tare the scale.3. Once the scale has been tared, the scale will display the weight of a previou
Instruction Manual - TORBAL AGC Series5. Place the sample on the pan. The weight of the sample will be displayed.Note: The reference sample must be eq
Instruction Manual - TORBAL AGC Series7. After setting the sample weight, the command READY will be displayed. Remove the sample and place the object
Instruction Manual - TORBAL AGC Series(F) TotalizingTotalizing is used to add and total weighing results. To use this function, follow the steps below
Instruction Manual - TORBAL AGC SeriesCHAPTER 4.PARTS DESCRIPTIONFront View Rear ViewPan Pan Base AC AdapterRS232 Serial Port Service Switch
Instruction Manual - TORBAL AGC SeriesManual When set to Manual, the weight of the object must be added manually. Once the weight stabilizes and the s
Instruction Manual - TORBAL AGC Series4. After the scale has been tared, Steps 1 through 3 can be repeated to add more weighing results to the grand t
Instruction Manual - TORBAL AGC SeriesCHAPTER 9.TECHNICAL INFORMATION72
Instruction Manual - TORBAL AGC SeriesCHAPTER 10.COMMON ERRORS AND TROUBLESHOOTINGError or Indicator Cause Explanation / Solution - - - - Below zero
Instruction Manual - TORBAL AGC SeriesCHAPTER 11.MAINTENANCECleaning and maintaining your AGC Precision Scale: Before cleaning the scale always unpl
Instruction Manual - TORBAL AGC SeriesCHAPTER 13.REPLACEMENT PARTSDescription Part No.A/C Adapter AGC9150Pan Assembly AGC9160RS232 Printer Cable AGC9
Instruction Manual - TORBAL AGC SeriesCHAPTER 5.UNPACKING THE SCALE AND GETTING STARTED1. Carefully remove the scale, pan, and all of its components
Instruction Manual - TORBAL AGC Series3. Once the pan has been installed, level the scale by adjusting the rear feet until the level indicator shows t